Testing & Quality Methodology
Analytical methods answer different questions, and the applicable test panel can vary by product and batch record. This page explains common methods that appear in research-material documentation and the limits of each method. It does not claim that every Core Research product receives the same tests or numerical release criteria.
Identity: mass spectrometry
Electrospray ionisation mass spectrometry can support molecular-identity assessment by comparing observed ions or a deconvoluted mass with the expected chemical species. The interpretation depends on the exact peptide or compound form, charge state, adducts and terminal modifications.
Mass spectrometry answers a molecular-mass/identity question. It does not by itself establish chromatographic purity, total content, sterility or suitability for a particular experiment.
Chromatographic composition: RP-HPLC
Reverse-phase HPLC separates components under a defined method and reports chromatographic signals, commonly as relative peak areas. A reported area percentage describes the detected composition under that method; it is not automatically the absolute mass fraction or measured peptide content.
Detection wavelength, column, mobile phase, gradient and integration rules matter. Use the HPLC purity guide for the detailed interpretation.
Water or moisture: Karl Fischer where applicable
Karl Fischer titration measures water. Where a product/batch record includes it, the result can be relevant to dry-material mass basis and stability assessment. It does not establish peptide identity or chromatographic composition.
Appearance
A visual observation can record colour, cake/powder form, clarity or visible damage where those attributes are relevant. Appearance is a receiving or specification field, not a substitute for molecular identity or compositional testing.
Specification, result and release are different concepts
A specification is an acceptance criterion defined for a material or process. A result is what a stated method measured. A release decision compares the applicable evidence with the applicable criteria. Those concepts should remain separate in a product or batch record.
Do not infer a universal ≥99% purity threshold, endotoxin limit, water-content limit or mass-spectrometry panel from this methodology page. The relevant product/specification and batch document own those values when they are actually documented.
Four checks before relying on a batch report
| Question | Why it matters |
|---|---|
| Which exact material and lot? | A report for a different batch, salt or blend cannot establish the material in the container. |
| Which test and measurement? | Identity, chromatographic composition, quantitative content, moisture and microbiological attributes are separate questions. |
| What result and acceptance criterion? | A numerical result should not be confused with an unverified advertised specification. |
| Who issued the document, and when? | The issuer, report identifier, date and applicable method help establish traceability. |
Ask for the documentation actually available for the relevant product and lot; the presence of this general methodology guide is not proof that every test has been performed on every Core Research batch.
How Core Research presents evidence
Product pages identify the catalogue material and supplied presentation. The Batch Documentation page explains how to request or match available records for a product or received lot. The Supplier Verification page explains the difference between supplier records, analytical results and accreditation claims.
Research-use boundary
Analytical documentation describes material evidence; it does not convert a Research Use Only material into a medicinal product or establish human/veterinary suitability. Core Research products are supplied for laboratory research only.